HELP: Advanced Database Error Being Displayed When Capturing Digital X-rays

 

SYMPTOMS

When trying to acquire a sensor image in a version of XrayVision before v3.5 b44, you may receive a "advanceddatabase.dll could not be mounted" error..

 

CAUSE

This error is a result of a sensor system not being selected before the user tries to do a quick sensor capture or tries to capture an image without first selecting the digital x-ray system.  Before build 44, even if the sensor system appeared to be selected, the software previously did not actually select the system until the user physically clicked on a system.  

 

RESOLUTION

This occurrence has been corrected in build 44 and later of the software.  To fix this error in versions before build 44, simply go into regular sensor capture screen and physically click on one of the system systems.  Once the user has performed this step once, they do not need to repeat it each time they run the software.